TECHNICAL DETAIL
A glass column having Dia. : 25.4 mm,
Length 900mm made of Borosilicate Glass.
Water Circulation by FHP Pump, Crompton/ Sharp make from water tank made of Stainless steel 304 grade, 1.2 mm thick, capacity 60 liters.
Air and water flow rate measured by using flowsensors with local digital display.
Pressure drop measurement one by using Differential pressure transmitter.
Buffer tank with safety valve & pressure gauge and separator made of Stainless steel 304 grade.
All the ball Valves should be imported. Operating temperature range should be -40°F to 302°F, Working pressure of valve should be in range of 600psi & valves use in vacuum pressure up to 2x10-5 psi.
High speed Data Acquisition system, model NI-USB 6001, National Instruments Make with 16 Inputs, Multifunction I/O, compact DAQ card.
The whole set-up well designed and arranged on a rigid structure painted with industrial PU Paint |